Contributors

Graham Speight, Ezam Uddin, Aysel Heckel, Venu Pullabhatla, James Reid and Lyudmila Georgieva

 

Introduction

A wide variety of chromosomal abnormalities are associated with Chronic Lymphocytic Leukaemia (CLL). Currently, comprehensive genetic research into CLL requires multiple testing strategies with high associated costs.

Somatic mutations (Single Nucleotide Variants (SNVs) and insertion/deletions (indels)) can be identified by next generation sequencing (NGS), but copy number alterations (CNAs) currently require additional cytogenetic methods including karyotyping, fluorescence in situ hybridization (FISH) and microarrays. For a more complete picture, follow-up assays are required to determine the trigger(s) behind the malignant transformation and to characterize the genetic profile to aid research into prognosis and disease management.

In this study, we tested the capability of SureSeq™ NGS panel to overcome the challenges with detecting CNAs currently experienced and provide a possible future single test to be developed for CLL.

 

Methods

The SureSeq hybridization-based approach was used throughout this study; the workflow of this is outlined in Figure 1.

Figure 1: OGT SureSeq workflow.Figure 1: OGT SureSeq workflow. The SureSeq workflow allows users to go from extracted DNA to sequencer in 1.5 days with minimal handling time. 

We utilized a SureSeq CLL CNV - 14 gene panel and associated library preparation kit to determine whether this approach can be used for detection of somatic CNAs as well as SNVs and indels.

CLL CNV - 14 gene panel can be used for:

  1. Detection of SNV and indels in 14 genes - ATM, PLCG2, BIRC3, BRAF, TP53, XPO1, SF3B1, KRAS, MYD88, SAMHD1, NOTCH1, BTK, CXCR4 and SRY
  2. Detection of somatic CNAs within 5 chromosomal regions – 17p (covering TP53), 11q (covering ATM), 13q (covering RB1/DLEU2/DLEU7), 6q (covering MYB) and Trisomy 12

We used a hybridization-based enrichment approach for library preparation and analysed 15 research samples* with known CNAs. We also analysed 24 samples (Coriell, Camden, NJ) with no known CNVs in the regions of interest which were used as control/reference samples. The resulting libraries were sequenced using a 2x150 bp read length protocol on an Illumina MiSeq®. All research samples were also processed on the Cytoscan™ HD microarray and associated software (Affymetrix®) allowing for comparison of findings of copy number alterations in each sample.

 

Bioinformatics analysis

Data sequencing analysis including CNA detection was performed using Interpret, OGT’s complimentary gene variants and CNV detection software.

 

Results I

Confident detection of SNV and indels in 14 genes

We achieved high depth (>2000x) and excellent uniformity of coverage across the targeted 14 genes which enabled the confident detection of low frequency gene specific SNVs and indels.

Figure 2: Example of SF3B1 exon 15 hotspot variant Lys700Glu with frequency 4.8%.Figure 2: Example of SF3B1 exon 15 hotspot variant Lys700Glu with frequency 4.8%. Data generated with OGT SureSeq protocol averaging ~2000x deduplicated coverage. Depth of coverage per base (grey).

Figure 3: Example of TP53 exon 4 frameshift deletion (TP53 c.124del) with frequency 38.9%.Figure 3: Example of TP53 exon 4 frameshift deletion (TP53 c.124del) with frequency 38.9%. Data generated with OGT SureSeq protocol averaging ~2000x deduplicated coverage. Depth of coverage per base (grey).

 

Validation of the CLL CNV - 14 gene panel and enhanced CNA detection software with 15 research samples
  • Data presented here are from 15 research samples (1 control with no CNAs in the regions of interest) that were processed using the OGT workflow in combination with OGT’s complimentary gene variant and CNV detection software.
  • Table 1 details the range of CNAs detected from 15 research samples. These include CNAs on chromosomes 17p, 11q and 13q. No CNAs were identified in the control sample.
  • CNA events were reported in samples with predicted tumor content as low as 25%.

Table 1: Data generated with the CLL CNV - 14 gene panelTable 1: Data generated with the CLL CNV - 14 gene panel using a combination of OGT workflow and enhanced CNV detection software was 100% concordant with independent findings (West Midlands Regional Genetics Laboratory – Birmingham, UK).

 

Results II

Confident detection of Copy Number Alterations

Using the OGT workflow we were able to reliably detect somatic copy number alterations in 15 research samples. For all samples, predicted CNAs were found to be 100% concordant with the reported events with the array data.

Figure 4: 181kb biallelic deletion within 13q14.2 including RB1.Figure 4: 181kb biallelic deletion within 13q14.2 including RB1.

Figure 5: 17.2 Mb deletion of 17pter to p11.2.Figure 5: 17.2 Mb deletion of 17pter to p11.2.

Figure 6: 11 Mb deletion of 13q14.2q14.3, including DLEU2 and DLEU7 genesFigure 6: 11 Mb deletion of 13q14.2q14.3, including DLEU2 and DLEU7 genes.

Figure 7: 56 Mb deletion of 11q14.1 to 11q25Figure 7: 56 Mb deletion of 11q14.1 to 11q25.

 

Conclusions

  • Superior uniformity of coverage from a hybridization-based enrichment using the SureSeq CLL CNV - 14 gene panel allowed simultaneous detection of SNVs, indels as well as larger structural alterations in a single assay.
  • We have demonstrated the capability of a SureSeq CLL CNV - 14 gene panel to detect complex rearrangements, ranging from a single gene (10 kb deletion covering TP53) to whole arm somatic deletions in samples with tumor content as low as 25%.
  • Our approach allows for the simultaneous evaluation of numerous chromosomal and gene-specific aberrations using a single assay.

 

Acknowledgements

*Samples kindly provided by West Midlands Regional Genetics Laboratory - Birmingham, UK.

 

SureSeq™: For Research Use Only; Not for Diagnostic Procedures.

Related content

  • Share
You might also be interested in

Assessment of Interpret software on low-frequency variants using reference standards

  • Resource type: Scientific poster
  • Application: Haematology, Solid tumour
Read

Saving time and costs in FISH automation

  • Resource type: Customer interview
  • Application: Haematology, Solid tumour
Read
CTA Icon

Stay up-to-date with the latest news from OGT, including new products, support resources, and our DNA Dispatch newsletter