Jacqueline Chan, Sabine Eckert, Lyudmila Georgieva and Graham Speight
OGT, Begbroke Science Park, Begbroke Hill, Woodstock Road, Begbroke, Oxford, UK
One of the challenges in cancer research is the high level of genetic complexity and tumor heterogeneity.
Research that generates detailed information about the genetic profile of each individual tumor will further our understanding and may be used in the future to guide treatment strategies1 .
NGS has enabled the simultaneous study of multiple mutations in high-penetrance cancer predisposition genes. However, tissue biopsies are typically archived as formalin-fixed, paraffin embedded (FFPE) blocks which can significantly compromise the quality and amount of nucleic acids available for genomics research.
To overcome these issues, we have used the SureSeq™ FFPE DNA Repair Mix, in combination with a hybridization-based NGS custom enrichment panel, the SureSeq Ovarian Cancer Panel (Table 1) to identify somatic variation in key DNA repair genes associated with ovarian cancer.
Table 1: Key ovarian cancer-related genes in the SureSeq Ovarian Cancer Panel.
To evaluate the application of a hybridization-based approach we:
The SureSeq hybridization-based enrichment was used throughout this study; the workflow of this is outlined below in Figure 1.
Figure 1: OGT SureSeq workflow. The SureSeq workflow allows users to go from extracted DNA to sequencer in 1.5 days with minimal handling time.
We tested a range of FFPE-derived DNA and formalin compromised DNA (Horizon Diagnostics - HD803 and HD799) and found pre-treatment with the SureSeq FFPE DNA Repair Mix significantly improves the number of on-target reads, thereby increasing the flexibility of the assay (Figure 2A). Use of the Repair mix also enables a reduced DNA input down to 50 ng to be used (if necessary) whilst maintaining a good depth of coverage (Figure 2B).
Figure 2: Example data obtained using FFPE DNA extracted from colon cancer samples and Horizon Diagnostics Quantitative Multiplex Formalin Compromised Reference Standards. Panel A shows that the SureSeq FFPE DNA Repair Mix improves on-target rate; Panel B demonstrates the Repair mix permits the use of lower DNA inputs whilst maintaining depth of coverage.
To confidently call low frequency variants, NGS reads need to be evenly distributed across all regions of interest. Uniformity of coverage is a useful value with which to compare this distribution and can be expressed as the percentage of target bases that have >20% of the mean coverage.
As reported extensively in the literature1-3, we found the uniformity of coverage from hybridization-based capture approaches, such as SureSeq, consistently outperform those enriched using amplicon-based methods (Figure 3). The uniformity of coverage for most samples is >99% of bases covered at >20% of the mean, ensuring that all bases within the panel can be assessed confidently. In addition, the use of hybridization-based capture instead of amplification-based enrichment allows the removal of PCR duplicates which can obscure the minor alleles present within a sample.
Uniformity of coverage using amplicon and hybridization-based approaches
Figure 3: Assessment of the uniformity of sequencing coverage from FFPE-derived DNA using an amplicon and the SureSeq hybridization-based capture approaches. Enrichment by SureSeq (dark blue bars) demonstrates better uniformity than that of an amplicon-based approach (green bars). Samples are ordered by increasing DIN determined by Agilent 2200 TapeStation – value in brackets.
We tested over 100 EOC samples determined by pathology to contain ≥40% tumor cells and identified one or more deleterious TP53 variant(s) with the minor allele frequencies (MAF) ranging from 1 to 87%. In addition to the mutations in TP53, some samples were found to have variants in BRCA1 (Figure 5). Figures 4 – 5 were visualized using Integrated Genomics Viewer4 ; the grey vertical bars denote the depth of coverage per base, green horizontal bars the targeted region, and the red heatmap - the GC content.
Figure 4: This sample (DIN score 3.0) was found to have a single base insertion - Cys124TrpfsTer25, at 2.8% in exon 10 of TP53 (transcript NM_ 000546) which results in a premature termination. It also contains a germline SNP (rs1042522).
Figure 5: TP53 exon 6 (panel A) and BRCA1 exon 16 (panel B). This sample (DIN score 2.6) contains an 18 bp deletion in TP53 at 23% and a 5 bp deletion of 28% allele frequency in BRCA1.
Samples kindly provided by –
aProf. Charlie Gourley (Cancer Research UK Edinburgh Centre)
bProf. Rober Zeillinger and Dr. Nicole Concin (Medical University of Vienna and Medical University, Dept. of Gynecology and Obstetrics, Vienna, Austria)
SureSeq: For Research Use Only; Not for Diagnostic Procedures.